Cautions for operation of moisture analyzer

Serial number

Precautions for operation of Guanya Moisture Analyzer

One,

Weighing instability occurs:

â‘ 

The bottom protection screw is not loose or not in place (generally 3-4mm) ※ More common

The sample holder is not placed properly

Environmental impact

â‘¡

â‘¢

two

light is not on:

â‘ 

There is no “↑” head, (the “↑” should be pressed after putting the sample, the heater will be on when the test button is pressed)

The sample weight must be greater than 0.5g.

The heater is not tightened and is loose.

The filament is broken.

â‘¡

â‘¢

â‘£

three

There is a large test error:

â‘ 

Each sample volume should be consistent

â‘¡

Set the temperature to be consistent

â‘¢

Test start temperature must be consistent

â‘£

Whether the sample is uniform.

⑤

Operating environment has an impact

â‘¥

Low voltage (this is often more common and can be solved with a voltage regulator)

four

The instrument is not powered:

â‘ 

The plug is not inserted properly

â‘¡

The fuse is blown. ※ (caused by voltage instability, more common)

Remarks

In order to avoid damage to the weighing system of the instrument, please tighten the screws of the bottom protection weighing system during the process of moving or transporting the instrument. If the instrument will not be used for a long time, please close the small cabin door on the right, and then use the original packaging materials to avoid damage.